Technical Parameter
SpO2
1.Measurement range: 0-100%
2.Accuracy: +-2% during 70%-100%
+-3% during 60%-69%
Pulse Rate
3.Measurement range: 30bpm-250bpm%
4. Accuracy: 1 bpm or +-2%. Take the bigger one
Visual Displays and Indicators:
5. SpO2 % display: Red LED. Variable brightness
6. Pulse Rate display: Green LED, variable brightness
7. Pulse intensity: Red LED 8 bars
8. Power Indicator Light: Green; Normal
Low Power voltage: Red and Blink
9.Low perfusion indicator: Red bar
Audio Indicator:
1.Probe off audio indicator (every thirty seconds)
2.Pulse audio indicator
3.Key-pressing indicator
4.Low power voltage
Environmental Specifications:
1. Temperature Range: Operating: (0 -- +50)C
2.Transport/Storage: ( -40 -- +70)C
3. Relative Humidity: Operating: 10-95%
4.(noncondensing) Transport/Storage: 0-95%
5. Vibration: Meets or exceeds ASTM 4728
(Method C) MIL-STD- 810E.
(Method 514.4, Section 1.3.4.3) IEC 68-2-37
6.Drop/Shock. Meets or exceeds IEC 68-2-32
(Procedure 1) MIL-STD- 810E.
(Method 516.4, Section 1-3.6, Procedure IV)
Power Requirements:
1.Four standard ''AA'' size alkaline batteries
Battery Life:
2. Twelve Hours
Safety Standards:
1. EN60601-1 (IEC-601-1)
2.GB 9706-1
Regulation Complied:
1. Quality System: ISO 9001 ISO13485
Dimensions:
2.24cm*16cm*7cm
Net Weight (with batteries / without batteries)
3.195g/100g
Gross weight: 480g
Package:
1. 1 set into one box in size 24cm*16cm*7cm and gross weight 1kg;
2. 5 sets to one carton in size 37cm*25cm*18cm and gross weight 5kg;
3.10 sets to one carton in size 48cm*37cm*18cm and gross weight 8kg.
4.20 sets to one carton in size 37cm*35cm*48cm and gross weight 15kg