Microplate Reader Specifications
Technical Specifications
(1) Range of working wavelength is from 400nm to 700nm. It has 4 filters when leaving factory: 405nm, 450nm, 492nm, 630nm;
(2) Repeatability: <=1.0%;
(3) Stability: <=?.005A/10min;
(4) Detection modes: single wavelength or dual wavelengths;
(5) Display: 240x128 pixels LCD screen;
(6) Print: external printer;
(7) Data Link: RS232 serial port;
(8) Output: absorbance, qualitative judgment, value of P/CO, concentration, judgment of normal value and quality control value
(9) Dimensions: 420x290x180mm:
(10) Net Weight: 10Kg.
(11) Gross Weight: 14Kg.
Microplate Washer Specifications
3.1 You can wash the 96-holes enzymatic plate by means of washing 8 or 12 holes every line.
3.2 The number of lines for each time scouring can be adjusted from 1 to 12.
3.3 The time of an enzymatic plate steeping can be adjusted from 0 to 3600 seconds.
3.4 The mean error of the injected washing liquid volumes in the holes of enzymatic plate is no more than 5%.
3.5 Average residue of scouring liquid is no more than 2ul in each hole after the enzymatic plate is washed.
3.6 The automatically circulated scouring times can be adjusted from 1 to 20.
3.7 The instrument can save up to 100 programs.
3.8 The instrument can scour its pipeline automatically.
3.9 Dimensions: 480mm?60mm?50mm.
3.10 Net weight: 15Kg.
3.11 Gross weight: 18 Kg